Thermodiag
Press and hold the pairing button for 3 seconds: the blue LED will flash and then remain steady when the device is ready.
The blue LED flashes every second and the name of the Armband is displayed on the terminal screen. You can click on the name to pair the Armband.
Steady green LEDs indicate that the battery is full and the Armband is fully charged.
The green LEDs that flash every second indicate that charging is in progress. Leave the Armband on the Docking Station until the LEDs are steady green.
A full charge takes up to 6 hours after use.
Place the Armband in one of the designated slots on the Docking Station. The LEDs will flash to indicate charging.
The station indicator light turns solid green when the power supply is connected to the mains.
The Armband provides approximately 24 hours of continuous measurement before recharging.
Press and hold the power button for three seconds to launch the Thermodiag app.
Press and hold the same button for one second, select SWITCH OFF, then confirm.
Press START, complete the patient form, press the Armband pairing button (the LEDs will flash blue) and select it when it appears on the screen.
Press STOP, then confirm or export the data if necessary.
No, only use the power supply unit provided to avoid any electrical risk or damage.
Connect the USB-C cable to the docking station: push it in until it clicks, and then connect it to the monitoring tablet. A vibration confirms that charging has started.
The system operates between 5 and 40 degrees Celsius at 15% to 90% relative humidity without condensation.
Wipe the Armband with a cloth soaked in a 70% hydroalcoholic solution, then leave it to dry in the open air.
Yes, disinfect the monitoring tablet with the same solution after each shared use.
Insert the charged Armband into a new Bandtape as shown, then fit it onto the patient's arm.
After twenty-four hours, the adhesive loses its hold and may distort the data. Replace it.
The five-character number appears on the UDI label and near the optical sensor.
Keep the Armband within a radius of approximately 10 metres to ensure a reliable connection.
A green double-pulse LED indicates that acquisition is active but that the distance is excessive; move the devices closer together.
Battery below 20%, replace the Armband with a fully charged Armband.
Ensure that it is within ten metres. Press the pairing button on the Armband to restart pairing with the blue LEDs flashing, then restart the search on the monitoring tablet.
Check that charging is complete: green LEDs should be steady; if not, contact support to have the battery checked.
- Check the placement of the Armband on the patient's arm and ensure there are no tattoos or hair.
- Change the Bandtape.
- Use the other arm if necessary.
Yes, excessive hair growth interferes with sensor-skin contact. Shave the area or change arm.
Check the network coverage, move the monitoring tablet or restart it, then try again.
No, the use of the device during transport is prohibited in order to guarantee the reliability of the measurements.
Store the components in the case at temperatures between -10 and +50 degrees Celsius with 40% to 90% relative humidity.
Immediately unplug the charger, wait for it to cool down, and then contact support before using it again.
The IP67 standard tolerates splashes. Dry the Armband and continue use without immersion.
No, avoid immersion and remove the Armband before showering or bathing.
The model requires approximately one hour of continuous acquisition before displaying the core body temperature.
- On the monitoring tablet, click on the PATIENT button. Verify that the patient's physical characteristics entered into the software are correct: age, gender, height, weight.
- If the information is correct but the problem persists, stop using the system and restart it.
- Contact support if the problem persists.
Close and restart Thermodiag on the monitoring tablet to resolve a minor glitch.
Check the label affixed to the charger unit:
If it does not bear the marking CINCON TR18RDM-120 or if the voltage is different, do not use this equipment.
The device supports 15 to 90% relative humidity without condensation.
If the problem persists, contact F2D MEDICAL support to have the charging system inspected.
Thermodiag is available in a range of models to suit the needs of healthcare establishments: purchase, rental or specific partnerships. The terms and conditions depend on the context of use, the volume of equipment, and the type of service concerned.
Our team works with each establishment on a case-by-case basis to determine the most suitable solution.
Yes. A rental option can be considered, particularly for gradual rollout phases, pilot projects, or temporary needs.
Renting allows you to test the solution in real-world conditions without a heavy initial investment, while benefiting from F2D Medical's support.
Yes. F2D Medical can enter into partnerships with healthcare institutions for pilot projects, clinical evaluations, or usage studies.
These partnerships are built with medical and institutional teams, within a defined framework that complies with medical device regulations.
Pilot project: a test phase in real conditions, with a limited scope and duration, aimed at evaluating the clinical and organizational benefits of the solution.
Yes. F2D Medical works with hospital stakeholders (purchasing departments, management, biomedical engineering) to adapt to public and private sector purchasing processes.
The contractual terms and conditions are defined in accordance with the regulatory frameworks in force.
Depending on the package chosen, the offer may include:
- access to the Thermodiag device,
- implementation support,
- team training,
- technical support, and follow-up.
Simply contact us via the form on the website or by email.
Our team will get back to you to understand your context, constraints, and objectives in order to put together a tailored proposal.
Our team will get back to you to understand your context, constraints, and objectives in order to put together a tailored proposal.